922-20207001/1.0
95
Details
Device Properties
Tip
You do not have to enter in the Device Properties manually. Clicking
Discover
Sonar Heads
, then
Use Discovered Head
will automatically populate the Device
Properties fields.
•
Head Name
You can enter any name to identify the Sonar Head.
•
Serial Number
The unique serial number of the Sonar Head.
•
IP Address, Port, Subnet Mask, and Default Gateway
These network parameters are used to identify and locate the Sonar Head on
the local network. Your Sonar Processor must be on the same network as the
Sonar Head.
Tip
If you need to set up a unique network environment, you can change the IP
Address, Port, Subnet Mask, and Default Gateway of the Sonar Head using
the
Head Network Setup
dialog box.
•
Trigger Mode
If applicable, set the
Trigger mode
to synchronize the Sonar Head with another
acoustic device (such as a second Flexview Sonar Head or a sub-bottom
profiler). Set the trigger mode to
Master
to send out a sync pulse so that the
other acoustic device can be triggered. Set the trigger mode to
Slave
to be
triggered by the sync pulse from the other device. For the Flexview, the default
is set to
Master
.
•
Time Sync Mode
Your computer can keep accurate time by using a GPS or network time server
as the master time source. When connecting to the Sonar Head, it takes two
minutes to synchronize the time to within five milliseconds. The Flexview
software will then continuously keep the Sonar Head clock in sync for the
duration of the session. The status bar shows the time synchronization status.
Get Default
Click this button to use the default network settings to connect to a Sonar Head.
Autoconnect at program start
Check this box to connect to the Sonar Head automatically the next time the
software starts.