54
922-20207001/1.0
Observe that a confirmation dialog box opens.
d
Click
Yes
to start the upgrade.
Tip
If you click
Cancel
or
Stop
, the upgrade process will be aborted. The Sonar
Head software and firmware will not be changed.
e
Select
RX FPGA Configuration
(receive firmware) under
Configuration type
.
f
Click the file folder icon under
Configuration file
to browse and select the .RXF
file on your local drive.
g
Click
Write
, then click
Yes
to start the upgrade.
Note
This file may take a few minutes to program.
h
Select
TX FPGA Configuration
(transmit firmware) under
Configuration type
.
i
Click the file folder icon under
Configuration file
to browse and select the .TXF
file on your local drive.
j
Click
Write
, then click
Yes
to start the upgrade.
Note
This file may take a few minutes to program.
k
Close the
Head Firmware Configuration
dialog box.
5
Confirm the upgrade was successful.
a
Click
Setup
→
Disconnect
.
b
Disconnect the power to the Sonar Head, then power it up again.
c
Wait ten seconds for the software to discover the Sonar Head.
d
Click
Setup
→
Connect
to start the Sonar Head.
e
Click
Setup
→
Head Firmware Configuration
.
Observe that the
Head Firmware Configuration
dialog box opens.
f
Confirm that the software and firmware has been upgraded from the versions
you took note of earlier.
Flexview Reference Manual