7.5 Setting Instrument Constant
·
Refer to “24.9 Instrument Constant (K)” to get the instrument constant value. Set it as following:
PROCEDURE
Operating Procedure
Operation
Display
(1) From Status Mode Press CNFG to enter
config mode.
CNFG
设
Config
1. Obs. Condition
2. Instr. const
3. Date & time
4. Comms setup
5. Unit
⑵
Select“2. Instr. const”, press ENT (or press
numeric key 2)
2. Instr. const
+
ENT
Instr. const
:
1. V angle 0 point
2. Instr. const
3. Contrast
⑶
After selecting “2. Instr. const”, press ENT
(Press numeric key 2 is also ok) to enter
instrument constant setting screen.
The first row is
Additive constant.
The second row is
Multiplication constant.
“2.
Instrument
constant”
+
ENT
Instr. const
:
Constant
:
30 mm
Mul cons: 0 ppm
⑷
Enter the constant, press ENT, return to
instrument constant setting screen.
Enter
constant
+
ENT
Instr. const
:
1. V angle 0 point
2. Instr. const
3. Contrast ADJ
Note
:
The constant of the instrument has been strictly set in the factory, so generally the user needs not to set this
item. If through strict measurement (ex. in standard baseline field and by special measuring organization) it is
necessary, the user can do that.
The
Additive Contant
is effective under
Prism mode
only (measuring distance with prism).