KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Creating and Editing a Scan Recipe
Y Scan Size (
μ
m)
This parameter defines the size, in the Y-direction, of the 3D area to be scanned. It is
the area across which the number of scans defined in the parameter
Traces
are divided
up. (See
.)
Setting or Changing
Y Scan Size
- Use one of the following procedures:
Click the menu arrow to the right of the
Y Scan Size
field and click the
desired size.
Highlight the current number and type in the new number. (See also
Automatic Parameter Adjustment
.)
Multi-Scan Average
2D
The number of single identical scans which are performed and used to
create a scan data set that represents the average of the scans.
Spacing
3D
This is the distance between X scans performed across the Y direction of
the 3D scan area.
Scan Direction
Both
The direction in which the scan is performed.
Teach…
Both
Displays the
Teach Scan Length
screen that is used to determine the start,
center or end of the scan. Can also be used to drag a new scan length.
Show Position
Both
Displays the current position and provides an opportunity to set a new
position at which the scan, of scan length set in
X Scan Size
, is started, is
centered, or ends.
Table 3.14
3D Scan Parameters Summary (Continued)
Parameter Setting
2D, 3D or Both Description and Location
NOTE:
If the variable in the
Spacing
parameter is changed, the
Y Scan
Size
changes to accommodate the number of
Traces
at the new
Spacing
distance.
Figure 3.18
3D Scan Parameters
Set the
Y Scan
Size
by clicking
on the menu
arrow and
choosing the
desired size, or by
highlighting the
current number
and typing in the
new number.