0142530-000 AB
KLA-Tencor Confidential
13-23
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Stylus Change Procedure - Stylus Removal and Replacement
1.
Once the feature is chosen, choose
Accept Current Calibration Result
. (See
.)
2.
Click
Take Selected Action
. (See
.)
3.
After the scan calibration has been accepted, the
Calibrations
screen returns.
Close the Calibration screen.
Figure 13.35
Hand Selecting the Triangle Data in Analysis Screen
First
: Click on the
feature that represents
the triangle.
Second
: Click
on
Set=>
.
The Up and Down parameters are displayed
in the
Up Edge
and
Down Edge
fields
Figure 13.36
Accepting Adjusted Scan Results
Choose
Accept Current
Calibration Result
.
Click on
Take
Selected Action
to accept
the calibration results.