XY View Screen - Positioning the Scan Site
KLA-Tencor P-16+ / P-6 User’s Guide
4-14
KLA-Tencor Confidential
0142530-000 AB
3/13/09
shows the stage coordinate system (SEMI Standard M20-92) used by the
Profiler. The X and Y coordinates relative to the center of the measurement area are
displayed in the current stage coordinate area of the XY View window. The travel
area of the stage is limited to a circle 210 mm (8.2 in.) in diameter. (See
.)
Scan Site Positioning Procedure
4
1.
After the sample is loaded on the stage and the stage returned to the scan
position under the stylus, click
FOCUS
.
2.
Use one or more of the following methods to locate a scan site. (See
.)
Figure 4.9
Coordinate System of the KLA-Tencor Profiler Stage
The coordinates are P-16+
max coordinates.
P-6 is 75mm in each axis.
Table 4.10
Locating a Scan Site
Movement Required
Movement Method
To make a large move
across the sample
surface, use the
Sample
Navigation Window
.)
Sample Navigation Window
– The navigation circle
represents the stage area. Click the location on the
Sample Navigation Window to move to the
corresponding location on the sample. (See
)
Move to a different site in
the current
Video Display
Window
(See
Video Display Window
– Click the desired site in
.
)
The site moves so that the video crosshair are
centered on the chosen location.