KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Feature Find
Diagnostic Options
3
Listed below are options located in the Diagnostic options window.
Scan Options
3
This is a set of miscellaneous scan related options.
Linearity Calibration Only – Diagnostic Options
3
Table 3.28
Diagnostic Options
Option
Description
No Motion Scan
During the scan, data is collected but the stage does not
move.
NOTE
:
This scan is only available in 2D.
Do Not Null Before
Scan
No movement of the elevator (for nulling) occurs before
the scan is performed and the data collected.
Table 3.29
Standard - Diagnostic Options
Option
Description
No Back Scan Before
Scan
Back Scan is a technique where, immediately prior to
the scan, the stage moves the scan start position back
and begins the scan nulling and movement. The
mechanical portion of the system has an opportunity
to settle before actually reaching the beginning of the
data collection.
This option prevents the Back Scan positioning from
taking place.
No Noise Filter
This prevents postprocessing of the scan data with
cutoff filters.
No Leveling
This prevents postprocessing data leveling of scan
data.
No Linearity Correction
Only used during the Linearity Calibration.
Table 3.30
Linearity Calibration Only – Diagnostic Options
Option
Description
Use Raw Data
Raw data from the scan is presented with no
postprocessing; without scaling to the measurement
range.
NOTE:
This option has no useful application apart from
the Linearity Calibration.
No Stylus Arc
Correction
Data from the scan is presented with no postprocessing
arcal correction.
NOTE:
This option has no useful application apart from
the Linearity Calibration.