Stylus Change Procedure - Stylus Removal and Replacement
KLA-Tencor P-16+ / P-6 User’s Guide
13-6
KLA-Tencor Confidential
0142530-000 AB
3/13/09
8.
After
Yes
is chosen, another message box is displayed. (See
.) This
box states that the stylus can be changed.
Notice that the message box contains a caution telling the user to “ensure the
sensor is unlocked.” Disregard this part of the message, it is not for the
MicroHead II sensor assembly.
DO NOT CLICK OK
UNTIL THE STYLUS HAS BEEN CHANGED.
9.
Open the Stage door.
10.
Loosen the thumbscrew holding the stylus wrench to the side of the head and
slide the wrench out of its holder.
11.
The head of the stylus clamp screw is visible from the front of the instrument.
Place a finger under the Stylus Mount to support it while the screw is being
loosened. Loosen the screw by inserting the stylus wrench and turning the
wrench counterclockwise 1/2 turn. Be careful to apply turning torque only. Do
not push against the screw head any harder than is necessary to seat the wrench.
(See
.) Do not remove the screw.
Figure 13.6
Message Box for Stylus Change Permission
CAUTION
:
C
lick OK
ONLY WHEN
the new
stylus is installed.
CAUTION: Do not operate
the stage or any motor driven component
with the door open
or the system will have to be rebooted.
Figure 13.7
Supporting Stylus Mount During Stylus Change