0142530-000 AB
KLA-Tencor Confidential
4-29
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
XY View Screen - Aligning the Sample
Displaying Grid Numbers in the Die Grid Navigation
Window
4
From a
Teach Scan
/
Site
window, click
Die Grid
, then click
Display Numbers
.
(See
.)click
If the numbers are too small to see, increase the size of the Die Grid Navigation
Window by clicking and dragging the window’s vertical separator bar to the left.
To Change the Font and Color of the Grid Numbers
4
1.
Go to a
Teach Scan
/
Site
window, click
Die Grid
, click
Font
.
A standard font dialog box appears.
2.
Select the font attributes desired and click
OK
.
A
LIGNING
THE
S
AMPLE
4
Introduction
4
This procedure aligns the sample image with the X-axis of the view screen using a
straight feature on the sample. Two methods for accomplishing this, each of which
rotate the stage (theta movement) to accomplish the alignment, are detailed in the
following sections. With the sample features aligned with the X-axis, more accurate
scans can be taken and die grid navigation is more accurate.
Procedure
4
Aligning the Sample with the Instrument
4
This procedure assumes that the sample is already on the sample stage and ready for
alignment. The sample must have a straight, easily discernible feature that can be
used to aligned the sample features with the X-axis of the XY view screen.
1.
Click the
FOCUS
button in the tool bar. The stylus nulls on the sample surface
and the sample image comes into focus.
2.
Using the linear movement arrow buttons, locate the center of the feature to be
used for alignment.
3.
Use the arrow buttons to approximately center the screen crosshair in the center
of the feature. (Or, move the cursor to the center of the feature and click. This
should move the crosshair to that location.)
CAUTION:
It is very important that the chosen feature be such that it
lies in a straight line across the X-axis of the sample. A thin line is best
for use in the alignment procedure.
NOTE: Steps 4-11 apply to P-16+ only. For P-6, use manual theta
adjustment for stage.