U3022BH10
Operational Check
Operational Check
This section provides the test calibration procedure to confirm the U3022BH10 operational
performance with regard to Reflection Tracking. The Reflection Tracking limits provided
ensure that your test set is operating properly. The operational check is performed with the
system configured as in
without the U3020AH37. The PNA-X and the
Test Set are connected with the same eight semi-rigid cables as shown. The operational
check configuration does not apply to the high power devices and exclude PNA-X Ports 3
and 4. Ensure that the standard jumpers are installed.
Verification Limits
Specifications for the U3022BH10 Multiport Test Set are typical. System performance for the
PNA-X and Test Set are only characteristic and intended as non warranted information. A
functional certificate is provided for the U3022BH10 only.
It is recommended that you return your instrument to Keysight Technologies for servicing or
repair if the Test Set and PNA-X performance exceed the operational verification limits. Refer
to
"Contacting Keysight" on page 123
Table 21
14-Port Operation Verification Limit
1
1.
Reflection Tracking takes into account Source Loss,
Receiver Loss, Margin, and PNA-X Mixer Cal.
Frequency
Reflection Tracking
Ports 1 - 2
Reflection Tracking
Ports 3 or 12
10 MHz to 5 GHz
−
6 dB
−
4 dB
5 GHz to 10 GHz
−
7 dB
−
6 dB
10 GHz to 15 GHz
−
9 dB
−
8 dB
15 GHz to 20 GHz
−
12 dB
−
9 dB
20 GHz to 26.5 GHz
−
15 dB
−
11 dB
U3022BH10 User's and Service Guide