U3022BH10
Controlling the System with the PNA-X
PNA-X Multiport Mode
The PNA-X requires Option 551 for Multiport mode. The PNA-X must have the Test Set file
installed to enable the PNA-X to control the Test Set. The PNA-X Multiport mode allows you
to select measurement ports, complete an N-Port calibration, operate control lines, and
rename test ports. Refer to the Help menu for more information.
If you change from Multiport mode to Standard PNA mode, send command 0.0
and 16.0. This will default the source and receiver paths in the Test Set
and allow the PNA-X ports to operate. Refer to
Paths with Address and Data Inputs" on page
The <address> <data> commands will only function while the PNA is in
Standalone Mode. Select
Restart as a standalone PNA > OK
. Refer to
Figure
36
below.
Multiport Mode Operation
1. The Option 551 must be installed for multiport capability. To access the multiport
application select Utility > System > Configure > Multiport Capability... -OR-
(Instrument > Setup > External
H
ardware > Multiport Configuration...)
.
2. Select U3022BH10_PNAX_p4 (14-Port System), or U3022BH10_PNAX_x2_p4 (24-Port
System) from the drop-down menu. Select Restart as a multiport PNA with this testset
> OK. The PNA-X will restart the network application with the Test Set interface features.
If the Test Set is not available in the drop-down list, it will be necessary for you to copy the
required Test Set file to the PNA-X hard drive. The current version of the Test Set files are
available on the web at
http://na.support.keysight.com/multiport
. Copy the appropriate file
to
c:\Program Files (x86)\Agilent\Network Analyzer\testsets
directory.
Figure 36
Test Set Selection
46
U3022BH10 User's and Service Guide