U3022BH10
Operational Check
N5241/2A/B 1-Port Calibration and Verification Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque to 8
in-lb. For more information, refer to the Help menu. See
Figure 68
below.
2. Perform a 1-Port Calibration on Port 1. On the PNA-X, select
Response > Cal > Start Cal > Calibration Wizard -OR-
(Response > Cal > Other Cals)
.
a. If using a mechanical cal kit, select SmartCal (Guided calibration) > Next.
b. If using an ECal module, connect the ECal to a PNA-X USB port. Select Use Electronic
Calibration (ECal) > Next
3. Continue following the Cal Wizard prompts. In the
Select Cali
b
ration Ports an
d E
Cal Mo
d
ule
dialog box, click Clear All, then select Port 1 > Next > Measure.
Figure 68
1-Port Calibration
4. Continue to follow the prompts until the Calibration Completed dialog box appears.
5. At the Calibration Completed prompt, select Save As User Calset, type the name 999.1.
Overwrite the Cal Set if it already exists and Save.
If you do not have a key board, select Save As User Calset > Edit Name
and save as 999.x. X is the port number you are calibrating. See
Figure
69
below. Use the numeric keypad on the PNA-X front panel to enter "999.1."
Figure 69
Calibration Complete
Help
90
U3022BH10 User's and Service Guide