U3022BH10
Method 2 - Using the Test Set Address and Data Values
This method is available while the system is in Standalone mode
only.
The Address and Data values for the Test Set can be found in the
“Address and Data Values” section on page 75
The Test Set internal switch settings are programmed into the
firmware of the Test Set and can be controlled with the specific
Address and Data values. Each Address and Data value pair sets the
Source or Receiver switch paths. For an S-parameter measurement,
two pairs of Address and Data values will be needed.
Address and Data values are separated by a comma. Commands
should be separated by a new line, or carriage return.
For example:
CONT:EXT:TEST:DATA <address>,<data>
CONT:EXT:TEST:DATA 0,0
Example: CONT:EXT:TEST:DATA 0,0
Figure 51
Method 2 - Using Test Set Address and Data Values
Controlling the System with the PNA-X
60
U3022BH10 User's and Service Guide