Available Options
Test Set Options
The Test Set has one available Test Set option: Refer to
•
Standard - mechanical switches for low loss.
•
Option 129 - adds an external source path for Ports 1, 3, 5, 7, 9, and 11.
Mounting Kits
The Test Set available cable options:
•
U3022BH10-1CN Front Handle Kit 1CN105A* (5063-9229)
•
U3022BH10-1CM Rack Mount Kit without Handles 1CM130A* (5063-9216)
•
U3022BH10-1CP Rack Mount Kit with Handles 1CP126A* (5063-9236)
*
New dark color version (Keysight color palette 2015)
U3022BH10
Measurement Application Notes
Low Loss DUT Measurements
Low loss trace ripple is a product of switch repeatability and calibration technique. Low
insertion loss measurements exhibiting a 0.1 dB uncertainty ripple residing on the measurement
response is not uncommon.
Switch Configuration Command Speed
When writing address and data values to the test set directly, include a 15 ms wait before
issuing another address and data set. The delay of 15 ms is required for the test set internal
switches to settle. Failure to add the delay could cause the test set internal switches to stick or
cause intermittent failures.
Available Options
Table 2
PNA-X Configuration Requirements
14-Port Network Analyzer System
Options
Test Set File
System Figure
N5241/2A/B PNA-X
24-Port Network Analyzer System
N5241/2A/B PNA-X
400
400
U3022BH10_pnax_p4.tsx
U3022BH10_x2_pnax_p4.tsx
Network Analyzer Requirements
All network analyzers require Option 551 to be installed for multiport operation (N-Port error
correction and measurement capability).
Table 2
(below) provides a listing of other network
analyzer option requirements. In order for the measurement system to cover the full frequency
range of the Test Set option listed above, the network analyzer's frequency range should equal
or exceed that of the Test Set.
The Test Set files indicated in
Table 2
above must be installed into this network
analyzer file directory location: c:\Program Files
(x86)
\
Agilen
t\Network
Analyzer\testsets
12
U3022BH10 User's and Service Guide