U3022BH10
General Specifications
Power Requirements
Verify that the required ac power is available before installing the Test Set to the PNA-X.
•
100/120/220/240 VAC (50/60Hz)
•
The instruments can operate with mains supply voltage fluctuations up to ± 10% of the
nominal voltage.
•
Air conditioning equipment (or other motor–operated equipment) should not be placed on
the same AC line that powers the Test Set and PNA-X.
•
Table 3
below contains the maximum wattage for all instruments. This table can be use to
determine the electrical and cooling requirements.
This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be inserted only into a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor, inside or outside the instrument, is likely to make the
instrument dangerous. Intentional interruption is prohibited.
Table 3
Power Requirements
Standard Equipment
Instrument
Maximum
Wattage
N5241/2A/B
450
U3022BH10
350
14
U3022BH10 User's and Service Guide