U3022BH10
Controlling the System with the PNA-X
If the PNA-X has Option 029 or H29 (noise figure capability), verify that the
path configuration is set to Default. If the PNA-X is not in Default
configuration, all odd ports will not operate correctly. Select
Trace/Chan > Channel
>
Hardware Setup > Path Config
...
-OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
and in the
drop-down menu select
Default
>
OK
. Save this configuration as a User
Preset by selecting
Save
>
User Preset
>
Save current state as User
Preset
. Do not use the factory Preset (User Preset Off), the PNA-X will
return to Option 029 path configuration.
Figure 43
RF Path Configuration
1. On the PNA-X select Response > Cal Wizard -OR-
(Response > Cal > Other Cals...)
a. If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b. If using an ECal module, connect the ECal to an available PNA-X USB port and select
Use Electronic Calibration (ECal) > Next -OR-
(Response > Cal > Other Cals... > ECal...).
Figure 44
Begin Calibration
U3022BH10 User's and Service Guide
51