U3022AH10
System Setup, 14-Ports
5
. Install the two rear locking feet onto the U3022
B
H10. Looking at the front panel, the
N5242-20206 is the right foot and the
N5242-20207 is the left foot. Two
screws
(0515-2317) are included with
these rear feet
.
Figure
11
Install Locking Feet on U3022
B
H10
6
. Place the network analyzer on top of the Test Set and ensure that the front frame of the network
analyzer is positioned slightly forward of the
Lock-Links (shown in
Figure 12
below)
that are
attached to the Test Set. Slide the
network analyzer back so the locks engage the front frame
of the analyzer.
Figure
1
2
Locking the Analyzer to the Test Set
Test Set
Front Frame
Network Analyzer
Front Frame
Slide the network Analyzer
back to engage the lock link
in the back of the network
analyzer's front frame
Lock Link
screw
Preparing the Test Set
4
. Install the
Lock-Link clips (1600-1423) with screws (0515-1402) on the Test Set front frame
as shown in
Figure 10
.
Figure
10
Install Lock
-Link Clips
24