U3022BH10
Operational Check
12. Compare the Reflection Tracking (1,1) trace to the appropriate limits
(see
"Verification Limits," page 87)
. This can be done using Limit Lines (press Marker/
Analysis > Analysis > Limit Test). The trace should be above the limit. PASS will be
displayed if the limit lines are used.
13. Repeat
step 12
for Cal Sets "999.1" thru "999.12".
Figure 74
Example: 14-Port System Reflection Tracking Trace (Port 1)
Figure 75
Example: 14-Port System Reflection Tracking Trace (Port 11)
U3022BH10 User's and Service Guide
93