U3022BH10
Operational Check
9. From the Cal Set drop-down menu, select 999.1 and select Enable. Select the Reflection Tracking
(x,x), where x,x is the port being tested, term in the center drop-down menu and ensure that the
Enable and Error Terms check boxes are selected.
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits (see
). You may also create a table on the PNA-X and enter the limit line stimulus and response
values. Select Marker/Analysis > Analysis > Limit Test
> Limit Test ON and Limit Line ON > Global
pass/fail display ON, then click Show Table. The trace should be above the limit. PASS will be
displayed on the screen if the limit lines are used.
11.Repeat
step 9
and
step 10
for the Cal Sets that you have saved.
Figure 72
PNA-X Setting the Test Limits
Figure 73
Reflection Tracking Trace
Enable
Error Term
999.1
Limit Table
92
U3022BH10 User's and Service Guide