General Specifications
U3022BH10
System Reflection Tracking Performance
Reflection tracking takes into account the following system related parameters: Source Loss,
Receiver Loss, Margin and PNA-X Mixer Cal. The reflection tracking values given are typical and are
intended as non-warrantied information.
Table 6
Typical System Reflection Tracking Values
14-Port System, 10 MHz - 26.5 GHz (
10 MHz to 2 GHz
−
3
−
2
2 GHz to 5 GHz
−
3.5
5 GHz to 10 GHz
−
5
10 GHz to 15 GHz
−
8
−
6
15 GHz to 20 GHz
−
11
−
7.5
20 GHz to 26.5 GHz
−
14
−
10
Typical values are based on the measurements of two to four instruments.
−
6
−
4.5
Frequency
Ports 1 to 2
Ports 3 to 12
14-Port System, 10 MHz - 18 GHz (
10 MHz to 2 GHz
−
11.5
−
5
2 GHz to 5 GHz
−
8
5 GHz to 10 GHz
−
11
10 GHz to 15 GHz
−
22
−
14
15 GHz to 18 GHz
−
24.5
−
19
−
18
−
14
Frequency
Port 1
Port 2
−
10
−
2.5
−
5
−
7
−
17
−
11
−
19
−
12
−
15
−
12
Ports 3-11 (Odd Ports)
Ports 4-12 (Even Ports)
24-Port System, 10 MHz - 18 GHz (
10 MHz to 2 GHz
−
11.5
−
5
2 GHz to 5 GHz
−
8
5 GHz to 10 GHz
−
11
10 GHz to 15 GHz
−
22
−
14
15 GHz to 18 GHz
−
24.5
−
19
−
18
−
14
Frequency
Port 1
Port 2
−
13
−
17
−
21
−
27
−
32
Port 3
Port 4
10 MHz to 2 GHz
−
10
−
3
2 GHz to 5 GHz
−
5
5 GHz to 10 GHz
−
7.5
10 GHz to 15 GHz
−
17
−
9
15 GHz to 18 GHz
−
19
−
11
−
15
−
12
Test Set#1 (Odd Ports)
−
3.5
−
10
−
13
−
16
−
12
−
19
−
15
−
21
−
9
−
6
+19
+15
+10
+6
−
4
Test Set#1 (Even Ports) Test Set#2 (Odd Ports) Test Set#2 (Even Ports)
1
1. PNA-X power level set to -15 dBm for the measurement of this E-Term. The Port 3 path through the U3020AH37 has a limited, 700 MHz to
18 GHz, frequency range.
U3022BH10 User's and Service Guide
17