Interface Cable Connection, 14-Port System
s
the
interface cable
configuration of the U3022
B
H10
Multiport Test Set and how it should be configured to the N5242A
/B
Option 400 (4-Port)
PNA-X Network Analyzer.
1.
Remove the four front panel jumpers on the PNA-X (Ports 1 & 2) from the SOURCE and RCVR
ports. The Reference jumpers must remain installed.
2. If not already installed, connect the 16 front panel jumpers (N5222-20091), as shown in
3.
Connect the RF cables from the N5242A
/B
Network Analyzer to the Test Set. Torque each
cable
to 8 in-lb. Refer to
System Setup, 14-Ports
Over
-
torque will cause damage to the Test Set and may cause connectors to
spin or become loose.
4.
Add the front panel port labels
provided (U3024-80001)
to the N5242A
/B
. Refer to
•
Port 13 over Port 3
•
Port 14 over Port 4
5.
Connect the cable (N4011-21002) from the PNA-X Test Set I/O to the U3022
B
H10 Test Set
I/O
on
the rear panel. See
6. Install the eight semi-rigid cables (please note that it might be necessary to reform the
semi-rigid cable slightly to fit in place). Tighten to 8 in-lbs torque.
U3022BH10 User's and Service Guide