U3022
B
H10
If the ports have different addresses, two separate address commands are needed.
Port 3 is the Source (with Noise Source Path and Port 4 is the Receiver.
Source Port 3 = address 0, data 9, and Receiver Port 4 = address 16, data 16.
Two separate commands must be sent. You may use the same dialog box. The entry
will be 0.9 in one command line, and 16.16 in the second line.
Example 4: (Option 129) Refer to
Address and Data Values for Option 129
Setting the Test Port Paths with Address and Data
for Option 129
and
information to set the internal switch paths of the Test Set.
Example 3: (Option 129) Refer to
If the ports have the same address, only one command is needed.
Port 3 is the Source and Port 7 is the Receiver.
Source Port 3 = address 0, data 9, and Receiver Port 7 = address 0, data 48.
The data values are added together, for an entry of 0.57.
For more information, refer to the PNA-X Help menu.
Address and Data Values
78
U3022BH10 User's and Service Guide