Capacitance Limits
The following table describes the maximum load capacitance that can be tolerated in a series
configuration. If you suspect that your load capacitance exceeds these values, you must install the
series diode as previously described.
1 kW Models
in series
Maximum load capacitance with 2 units
in series
2 kW Models
in series
Maximum load capacitance with 2 units
in series
N6950A/N7950A
381 µF
N6970A/N7970A
763 µF
N6951A/N7951A
94 µF
N6971A/N7971A
188 µF
N6952A/N7952A
23 µF
N6972A/N7972A
46 µF
N6953A/N7953A
11 µF
N6973A/N7973A
22 µF
N6954A/N7954A
6 µF
N6974A/N7974A
12µF
N6976A/N7976A
5.5µF
N6977A/N7977A
3µF
Additional Cautions When Using a Boost Supply in Series
The following cautions only apply to 40V, 60V, 80V, 120V and 160V N6900 models without relay
disconnect option 760 or 761.
To Prevent Possible Equipment Damage:
l
Do not power up the boost supply until after the APS is turned on and initialized.
l
Turn off the boost supply first; then turn off the APS unit.
The following figure illustrates how a boost supply is typically connected.
2 Installing the Instrument
82
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...