![Keysight Technologies N6900 Series Operating And Service Manual Download Page 231](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217231.webp)
FUNCtion Command
[SOURce:]FUNCtion CURRent|VOLTage
[SOURce:]FUNCtion?
Sets the output regulation - voltage priority or current priority. In voltage priority mode the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Parameter
Typical Return
CURRent|VOLTage, *RST VOLTage
CURR or VOLT
Sets the output regulation to current priority: FUNC CURR
6 SCPI Programming Reference
231
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...