POWer Query
[SOURce:]POWer:LIMit? [MIN|MAX]
Returns the output power limit of the instrument.
AC mains circuits rated at 100-120 VAC cannot supply enough current to power either the 1 kW or the
2 kW models when operated at their full rated output power. Use this query to return the maximum
output power available when connected to 100-120 VAC mains. When the maximum available power
limit is exceeded, the instrument turns off the output and sets the CP+ status bit.
Parameter
Typical Return
(none)
<power limit>
Query the power limit: POW:LIM?
Keysight N6900/N7900 Series Operating and Service Guide
254
6 SCPI Programming Reference