![Keysight Technologies N6900 Series Operating And Service Manual Download Page 259](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217259.webp)
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent:MINMax?
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage:MINMax?
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables MIN/MAX logging values: SENS:ELOG:FUNC:VOLT:MINM ON
SENSe:ELOG:PERiod <
value
>|MIN|MAX
SENSe:ELOG:PERiod? [MIN|MAX]
Sets the integration time of an Elog measurement. Although the absolute minimum logging period is
102.4 microseconds, the actual minimum varies as a function of the number of readings that are being
logged (see
).
Parameter
Typical Return
0.0001024 to 60, *RST MAX
<period>
Specifies a datalog period of 0.01 seconds: SENS:ELOG:PER 0.01
SENSe:FAULt:STATe 0|OFF|1|ON
SENSe:FAULt:STATe?
Enables or disables remote sense fault detection. This condition is annunciated by the SF status bit.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Disables remote sense fault detection: SENS:FAULt:STATe OFF
SENSe:SWEep:NPLCycles <
value
>|MIN|MAX
SENSe:SWEep:NPLCycles? [MIN|MAX]
Sets the measurement time in number of power line cycles. Increasing the number of power line
cycles reduces the measurement noise on current and voltage measurements. Changing the NPLC
changes the number of points and time interval setting. The number of points in 1 NPLC depends on
the line frequency (see
).
Parameter
Typical Return
0.0003072 to 1,258,290,000,000, *RST 1
<NPLC value>
Specifies 100 power line cycles: SENS:SWE:NPLC 100
6 SCPI Programming Reference
259
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...