Front Panel Display at a Glance
Voltage and current meas-
urements
Displays the actual output voltage and current
Operating mode
Indicates one of the following:
OFF = the output is off
CV = the output is in constant voltage mode
CC = the output is in constant current mode
CP+ = the output is disabled by the positive power limit
CP– = the output is disabled by the negative power limit
VL+ = the output is in positive voltage limit mode
CL+/– = the output is in positive or negative current limit mode
OV = the output is disabled by the over-voltage protection
OV– = the output is disabled by the negative over-voltage protection
OC = the output is disabled by the over-current protection
OT = the over-temperature protection has tripped
PF = the output is disabled by a power-fail condition
SF = a sense lead fault has been detected
P = the instrument has been paralleled with another instrument
Inh = the output is disabled by an external inhibit signal
Unr = the output is unregulated
Prot = the output is disabled by a protection condition on another unit
Edp = the output is disabled by excessive output dynamics protection
UProt = a user-defined protection signal has disabled the output
IPK+/– = the output is in positive or negative peak current limit
CSF = a current sharing fault has occurred
T-on = the output is turning on
T-off = the output is turning off
Voltage and current
settings
Displays the programmed voltage and current. These settings may not match the
measured output voltage or current. For example, in constant voltage operation, the
output current setting (limit) may be set to 1 A, but the actual (measured) output current
must be less than 1 A for the output to remain in constant voltage mode. If the Current limit
is reached, the output will no longer be operating in constant voltage mode, but will be in
current limit mode. In this case, the actual output voltage will now be less than the output
voltage setting.
Status area
Indicates the following remote interface activity:
Err = an error has occurred (press Error key to display error message)
Lan = the LAN is connected and has been configured
IO = there is activity on one of the remote interfaces
1 Quick Reference
18
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...