![Keysight Technologies N6900 Series Operating And Service Manual Download Page 211](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217211.webp)
ARB Subsystem
ARB commands program the constant-dwell arbitrary waveforms. Constant-dwell waveforms can
have up to 65,535 points assigned to them, with the same dwell time for each point.
[SOURce:]
ARB
:COUNt <
value
>|INFinity
Specifies the number of times the Arb repeats.
:CURRent
:CDWell
[:LEVel] <
value
>{,<
value
>}|
<Block>
Specifies the level of each point in the Arb.
:DWELl <
value
>
Specifies the dwell time of each point in the Arb.
:POINts?
Returns the number of points in the Arb.
:FUNCtion
:TYPE CURRent|VOLTage
Specifies either a voltage or current Arb.
:TERMinate
:LAST 0|OFF|1|ON
Selects the output setting after the Arb ends.
:VOLTage
:CDWell
[:LEVel] <
value
>{,<
value
>}|
<Block>
Specifies the level of each point in the Arb.
:DWELl <
value
>
Specifies the dwell time of each point in the Arb.
:POINts?
Returns the number of points in the Arb.
[SOURce:]ARB:COUNt <
value
>|MIN|MAX|INFinity
[SOURce:]ARB:COUNt? [MIN|MAX]
Specifies the number of times the Arb repeats. Use the INFinity parameter to repeat the Arb
continuously.
Parameter
Typical Return
1 – 256, *RST 1
<count>
Programs a repeat count of 10: ARB:COUN 10
[SOURce:]ARB:CURRent:CDWell[:LEVel] <
value
>{,<
value
>}|<
Block
>
[SOURce:]ARB:CURRent:CDWell[:LEVel]?
[SOURce:]ARB:VOLTage:CDWell[:LEVel] <
value
>{,<
value
>}|<
Block
>
[SOURce:]ARB:VOLTage:CDWell[:LEVel]?
Specifies the level of each point in the Arb. Values are specified in either amperes or volts. The
minimum and maximum values depend on the ratings of the unit.
6 SCPI Programming Reference
211
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...