Sharing Deviation for Units of Equal Power (either 1 kW or 2 kW)
The following table gives the Gain and Offset values for paralleled units of equal power:
Paralleled Units (NT)
Gain Error % (G)
Offset Error % (K)
Gain and Offset Equations
2
0.200
0.6
G = 0.4%((NT – 1)/NT)
K = 0.6%(NT – 1)
3
0.267
1.2
4
0.300
1.8
5
0.320
2.4
Example (60 A load current) You have three 1 kW, 40 V, 25 A units connected in parallel drawing a
total load current of 60 A. Using the gain and offset values from the above table (G=0.267%; K=1.2%),
the worst-case deviation from the ideal current sharing contribution of 20 A for an individual unit would
be as follows:
∆IOUT
(WORST_CASE)
=±G(IOUT
(IDEAL)
) ±K(IRATING)
∆IOUT
(WORST_CASE)
=±0.267%(20A) ±1.2%(25A)
∆IOUT
(WORST_CASE)
=±0.353A
Note that the percent of deviation from the ideal becomes greater for smaller output currents,
because the offset error predominates. This holds true all the way down to zero current. If, in the
above example, the paralleled units were drawing no current (0 A) , the worst case deviation would
still be:
∆IOUT
(WORST_CASE)
=±0.267%(0A) ±1.2%(25A)
∆IOUT
(WORST_CASE)
=±0.3A
Sharing Deviation for Units of Mixed Power (1 kW paralleled with 2 kW)
Note that calculating the current deviation in this procedure is a little more complicated than the
previous one, because the 1 kW and the 2 kW units each contribute a
different
amount of current
toward the total load current. Ideally, the 2 kW units contribute twice as much current as the
participating 1 kW units.
The following table gives the Gain and Offset values for up to five paralleled units of mixed power:
Paralleled Units (NT)
Gain Error % (G)
Offset Error % (K)
Gain and Offset Equations
2
0.267
0.40
G = 0.8%((NT – 1)/(1 + 2(NT – 1)))
K = 1.2%(NT((NT – 1.5)/(2NT – 1)))
3
0.320
1.08
4
0.343
1.71
5
0.356
2.33
Keysight N6900/N7900 Series Operating and Service Guide
183
4 Using the Advanced Power System
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...