OUTPut [:STATe] 0|OFF|1|ON
OUTPut[STATe]?
Enables or disables the output. The state of a disabled output is a condition of zero output voltage
and zero source current. If output and sense relays are installed, the relays will open when the output
is disabled and close when the output is enabled.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Turns the output off: OUTP OFF
l
When output is enabled, the front-panel status indicator changes from OFF to indicate the oper-
ating status of the instrument (CV, CC, etc.).
l
Separate delays can be programmed for the off-to-on and the on-to-off transition using
OUTPut:DELay:RISE and OUTput:DELay:FALL.
l
Because of internal circuit start-up procedures and any installed relay options, OUTPut ON may
take between 12 and 38 milliseconds to complete its function in voltage priority mode, and
between 14 and 46 milliseconds in current priority mode.
OUTPut[:STATe]:COUPle[:STATe] 0|OFF|1|ON
OUTPut[:STATe]:COUPle[:STATe]?
Enables or disables output coupling. Output coupling allows the outputs of multiple instruments to
turn on and off sequentially according to their specified OUTPut:DELay:RISE and OUTput:DELay:FALL
programming delays. This parameter is saved in non-volatile memory.
Parameter
Typical Return
0|OFF|1|ON
0 or 1
Turns the output coupling state on: OUTP:COUP ON
l
You must connect and configure the ONCouple and OFFCouple digital connector pins of all syn-
chronized instruments as described in the
section.
l
Because some power supplies have different minimum delay offsets, you must also specify a com-
mon delay offset for all of the synchronized units. This value must be the largest delay offset of the
synchronized group. Use OUTPut:COUPle:MAX:DOFFset? to query the delay offset for each unit.
The largest value returned must be specified as the common delay offset for each unit.
OUTPut[:STATe]:COUPle:DOFFset <
value
>|MIN|MAX
OUTPut[:STATe]:COUPle:DOFFset? [MIN|MAX]
Sets a delay offset to synchronize coupled output state changes. Units are in seconds. Setting this
time to the maximum delay offset specified for any instrument that is being coupled will cause all
coupled outputs to synchronize to the turn-on times specified by OUTPut:DELay:RISE. This parameter
is saved in non-volatile memory.
Keysight N6900/N7900 Series Operating and Service Guide
248
6 SCPI Programming Reference
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...