-144 Character data too long
The character data element contains more than 12 characters.
-148 Character data not allowed
A discrete parameter was received, but a string or numeric parameter was expected.
-150 String data error
Generic string data error
-151 Invalid string data
An invalid character string was received. Check that the string is enclosed in quotation marks.
-158 String data not allowed
A character string was received, but is not allowed for this command.
-160 Block data error
Generic block data error
-161 Invalid block data
The number of data bytes sent does not match the number of bytes specified in the header.
-168 Block data not allowed
Data was sent in arbitrary block format but is not allowed for this command.
-170 Expression error
Generic expression error
-171 Invalid expression data
The expression data element was invalid.
-178 Expression data not allowed
Expression data element was sent but is not allowed for this command.
Execution Errors (these errors set Standard Event Status register bit #4)
-200 Execution error
Generic syntax error
-220 Parameter error
A data element related error occurred.
-221 Settings conflict
A data element could not be executed because of the present instrument state.
6 SCPI Programming Reference
301
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...