![Keysight Technologies N6900 Series Operating And Service Manual Download Page 258](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217258.webp)
SENSe:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON
SENSe:CURRent[:DC]:RANGe:AUTO?
Enables or disables seamless measurement auto ranging.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Enable current measurement autoranging: SENS:CURR:RANG:AUTO ON
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer] <
value
>|MIN|MAX
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX]
Selects the Elog current measurement range. The value that you enter must be the highest value that
you expect to measure. Units are in amperes. The instrument selects the range with the best
resolution for the value entered.
Parameter
Typical Return
0 to maximum
<max current>
Selects the 1 A range: SENS:ELOG:CURR:RANG 1
l
The MAX query returns the maximum DC current that can be measured on the present range.
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO?
Enables or disables Elog seamless measurement autoranging.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Enable Elog current autoranging: SENS:ELOG:CURR:RANG:AUTO ON
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent?
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage?
Enables or disables the Elog current or voltage measurement function.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables datalog current measurements: SENS:ELOG:FUNC:CURR ON
Keysight N6900/N7900 Series Operating and Service Guide
258
6 SCPI Programming Reference
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...