Front Panel Menu Reference
SCPI Command
Not available
To enable voltage or current measurements:
SENS:ELOG:FUNC:VOLT ON
SENS:ELOG:FUNC:CURR ON
To enable min/max measurements:
SENS:ELOG:FUNC:VOLT:MINM ON
SENS:ELOG:FUNC:CURR:MINM ON
Keysight N7900 models have two current measurement ranges, a high and a low range (see
). A seamless current ranging function ensures that no data is lost due to range
switching. Seamless ranging is enabled by default. The commands to enable seamless current
measurement ranging are:
Front Panel Menu Reference
SCPI Command
Not available
To enable seamless elog autoranging:
Specify the Integration Period
The integration period can be set from a minimum of 102.4 microseconds to a maximum of 60 seconds.
Front Panel Menu Reference
SCPI Command
Not available
To set an integration period of 600 microseconds:
During the integration period, Elog samples are averaged, and the minimum and maximum values are
tracked. At the end of each integration period the average, minimum, and maximum values are added
to the internal FIFO buffer.
Although the absolute minimum integration period is 102.4 microseconds, the actual minimum
depends on the number of measurements that are being logged. The formula is 102.4_microseconds
X #_of_measurements. For example:
102.4 microseconds: 1 measurement (voltage or current)
204.8 microseconds: 2 measurements (voltage and current)
409.6 microseconds: 4 measurements (min+max+current)
If the specified integration period is at or near the minimum logging intervals, the data format must be
specified as binary. If the REAL format is not specified, the data will be in ASCII format and the
minimum logging intervals will typically be up to five times longer than what can be achieved with
binary format.
Front Panel Menu Reference
SCPI Command
Not available
To set the data format to REAL:
Keysight N6900/N7900 Series Operating and Service Guide
159
4 Using the Advanced Power System
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...