![Keysight Technologies N6900 Series Operating And Service Manual Download Page 308](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217308.webp)
Recommended Test Equipment and Setups
Test Equipment
The test equipment recommended for the performance verification and adjustment procedures is
listed below. If the exact instrument is not available, substitute calibration standards of equivalent
accuracy.
Instrument
Requirements
Recommended Model
Use*
Digital Multimeter
Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A
V, C
Current Shunt
15 A (0.1
Ω
) 0.04%, TC=5ppm/
Ω
C
100 A (0.01
Ω
) 0.04%, TC=5ppm/
Ω
C 300 A
(0.001
Ω
) 0.04%, TC=5ppm/
Ω
C
Guildline 9230/15
Guildline 9230/100
Guildline 9230/300
V, C
Electronic load
80 V, 200 A minimum
2 - Keysight N3300A mainframes; 6 -
Keysight N3305A modules
V
GPIB controller
Full GPIB capabilities
Keysight 82350B or equivalent
V, C
Oscilloscope
Sensitivity: 1 mV Bandwidth
Limit: 20 MHz
Probe: 1:1 with RF tip
Keysight Infiniium or equivalent
V
RMS Voltmeter
True RMS
Bandwidth: 20 MHz
Sensitivity: 100 µV
Rhode and Schwartz Model URE3 or
equivalent
V
Differential Amplifier
Bandwidth: 20 MHz
LeCroy 1855A, DA1850A, or equivalent
V
Terminations
1 – 50
Ω
BNC termination
2 – 50
Ω
, 1/8 W termination resistors
V
Power Supply
160 V, 200 A, 2kW
Keysight N8754A, N8758A, N8761A
V, C
* V=Verification C=Calibration
7 Verification and Calibration
308
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...