Keysight N5990A MIPI M-PHY Receiver Test User Guide
3
Calibrations and Test
Procedures
Example of MIPI M-PHY Calibration / Test Procedure
During the execution of all calibration and test procedures, the results are
displayed automatically in a data table as well as graphically. The viewer
can be either an
MS-Excel
or a
HTML
worksheet; this can be chosen in the
Station Configuration
). Once a specific calibration or test
procedure is finished, the
MS-Excel/HTML
worksheet is closed. To
re-open it at any time, double click on the respective procedure in the
respective tree.
To save calibration data worksheets in a workbook, go to
File > Save
Results as Workbook...
at any time. It is recommended you carry out this
step at least at the end of each N5990A run.
If the calibration and test procedures are conducted several times during
the same N5990A run, the result worksheets are combined in the
workbook.
If you conduct a test procedure without prior execution of calibration
procedures in the same test run, only the test results will be saved to the
workbook. As a safety feature, all calibration and test results are saved by
default to the N5990A “Tmp” directory (refer to the “Keysight N5990A Test
Automation Software Platform Installation Guide
”
). The sub-folder
Results/M-PHY Station contains the Excel files of the final results
measured at each calibration and test procedure. In addition to the
calibration data worksheets, the calibration data files are generated. These
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......