Keysight N5990A MIPI M-PHY Receiver Test User Guide
117
Calibrations
4
RJ Calibration
This calibration is available for all hardware configurations. It is only
defined for HS Gear3..
Figure 67
RJ Calibration
Purpose
The RJ calibration for HS-G3 combines the two procedures defined for
HS-G1/G2 "Short Term RJ Calibration" and "Low Frequency RJ". It
eliminates the Low Frequency RJ component, so that the entire RJ budget
of 0.17UI is allotted to wide-band RJ, above 10MHz.
This calibration should be run once at each HS-G3 data rate (see
). For spec 4.1. this calibration is only mandatory for B-series (for
A-series, the procedure is shown in the test tree as informative).