10
Keysight N5990A MIPI M-PHY Receiver Test User Guide
A
Appendix
Figure 3
DSGA UniPro Test Mode setup
• J-BERT DSGA: Works similar to
M8020A
+
DSGA
.
The training sequence needs to have the following structure:
• Send PACP Test Mode Request (
TestModeReq()
macro).
• Bring TX out of HIBERN8 (
SetReq(…)
macro, set register 2B).
• Configure Rx transmission Mode, Gear, etc.
• Configure Tx transmission Mode, Gear, etc.
• Transmit Test Pattern with Equipment (
TestDataFrame
macro).
• Request Frame and Error Counter (
GetReq(…)
macro, registers 15C0
and 15C1).
NOTE
In addition to the
M8020A
and the
DSGA
, a BIT-2100 Switch system is
required to alternate the DUT Rx input between the two generators.