174
Keysight N5990A MIPI M-PHY Receiver Test User Guide
5
HS Tests
Dependencies
The calibrations required for this test are as follows:
• Reference Clock Calibration
• Level Calibration Terminated
Results
An example of a HTML Viewer for the Test 2.1.5-HS-RX Differential
Termination Disable Time procedure is shown in
. The results
comprise of the following:
• The common parameter list (in
Expert Mode
)
• A test data table (Refer to
Figure 107
Example HTML Viewer for Test 2.1.5-HS-RX Differential Termination Dis-
able Time
Table 25
Test 2.1.5-HS-RX Differential Termination Disable Time data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the value is "Fail", the termination enable time is not with in the spec limit.
Termination Time Min
It is the shortest/minimum termination enable time in "ns" (nanosecond).
Termination UI Min
It is the shortest/minimum termination enable time in "UI".