Keysight N5990A MIPI M-PHY Receiver Test User Guide
35
N5990A M-PHY Station
2
of the instruments used in the M-PHY station require a VISA (Virtual
Instrument System Architecture) connection. To determine and validate
the VISA address, do the following:
1 Run the
VISA Connection Expert
. To access the VISA Connection
Expert, right-click the
Keysight IO Libraries Suite
icon in the task bar
and select
Connection Expert
.
2 Enter the instrument addresses in the
Station Configuration
wizard,
for example, by copying and pasting the address strings from the
Connection Expert
entries.
3 Click
Apply Address
before selecting the
Offline
check box to set
the instruments that are required to be online.
4 Click
Check Connections
to verify that the connections for the
instruments are established successfully.
If anything is wrong with the
Instrument Address
, a window is displayed
with a message describing the problem.