34
Keysight N5990A MIPI M-PHY Receiver Test User Guide
2
N5990A M-PHY Station
Oscilloscope Bandwidth
Auto
If the
Auto
option is selected, Keysight discovers the maximum supported
oscilloscope bandwidth automatically and adjusts to the recommended
value for each gear. Firmware version 5.20 or above is required for this
purpose.
Manual
If the oscilloscope firmware is below version 5.20, it can be complicated for
Keysight to find the maximum oscilloscope bandwidth automatically. To
avoid this situation, select the
Manual
option to set the desired bandwidth
manually.
Once the M-PHY station is configured, you must set the
Instrument
Address
. An example for the instrument configuration is as shown below
(see
Figure 8
M-PHY Instrument Configuration Window
After the installation process, all instruments are configured by default in
Offline
mode. In this simulation mode, the hardware does not need to be
physically connected to the test controller PC. N5990A cannot connect to
any instrument in this mode. In order to control the instruments that are
connected to the PC, the instrument address must be entered. The
address depends on the bus type used for the connection, for example,
GPIB (General Purpose Interface Bus) or LAN (Local Area Network). Most