Keysight N5990A MIPI M-PHY Receiver Test User Guide
11
Appendix
A
Unipro Script Generation
To operate in
UniPro
Test mode, it is necessary to replace the default
scripts by UniPro scripts. These are generated in the
UniPro Script
Generator
dialog.
From the
Configure DUT
dialog, go to
Default Sequence > Generate
UniPro Scripts
. The
UniPro Script Generator
dialog is displayed.
Figure 4
Unipro Script Generator Dialog
Here, you can select from several properties:
Test Pattern
You can choose between two test patterns supported by the UniPro
Adapter Layer:
• CJTPAT
• CRPAT
Tx Response
In
Receiver
tests, you can configure the DUT to transmit the response in
high or low speed:
PWM Gear 1
The DUT transmits the signal in low speed PWM. You must select this
option for setups that include the
DSGA
instrument as Error Detector
(
M8020A
+
DSGA
or
N4903B
+
DSGA
setups).