Keysight N5990A MIPI M-PHY Receiver Test User Guide
41
N5990A M-PHY Station
2
Expert Mode
Calibrations and tests can be conducted beyond the limits and constraints of the CTS; the test
parameters used in the calibration and test procedures are shown and can be modified by the user.
For Receiver and Transmitter Test Configuration
HS Gears
The following HS-Gears are available:
•
GEAR 1-A
•
GEAR 1-B
•
GEAR 2-A
•
GEAR 2-B
•
GEAR 3-A
•
GEAR 3-B
•
GEAR 4-A
•
GEAR 4-B
The selection of gears is not limited. If "x" number of gears are selected, the HS terminated and
non-terminated tests are available "x" times, in respective test sub-groups.
For receiver test, GEAR 3-A and 3-B are only available for spec. 3.10 and 4.10. GEAR 4-A
Nominal Data Rate
It sets the reference clock frequency to achieve the nominal Gear x-B data rate value as given in the
specification. In case of custom data rates, the reference clock is adjusted to achieve the HS custom data
rate.
Nominal Ref. Clock
For the selected reference clock frequency, the generated data rate is calculated as a multiple of the
reference clock.
Additional Gears
PWM Gear 1 is always tested. Additionally, PWM Gears from Gear 2 to Gear 7 can be selected.
The selection of gears is not limited. If "x" number of gears are selected, the PWM terminated and
non-terminated tests are available "x" times, in respective test sub-groups.
This option is only visible for PWM device type.
M-PHY Spec Version
Select the version using the drop down menu:
•
3.10
•
4.10
Ref Clock Frequency
The frequency can be chosen as follows:
•
19.2 MHz
•
26 MHz
•
38.4 MHz
•
52 MHz
Default Levels
Brings up a dialog where the voltage levels and the default values can be selected. Refer to Default
Levels for more details.
Default Timing
Brings up a dialog where the timing parameters (Prepare Length, Stall, Sleep, Sync Length) can be
defined. Refer to
for more details.
Default Sequences
Brings up a dialog where the HS, LS, and Squelch sequence files can be selected. Refer to
for more details.
Receiver Test Configuration
Parameter Name
Parameter Description
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......