Keysight N5990A MIPI M-PHY Receiver Test User Guide
43
N5990A M-PHY Station
2
Custom HS DR
Click
Custom HS DR
in the
Configure DUT
). A
window is displayed as shown in
. It allows custom data rates to
be added to the tested data rates list.
Figure 14
Custom HS Data Rate panel
Custom Data Rate
Enter the desired value into the
Custom Data Rate
text box and click
Add
.
The custom data rate
is added to the list of data rates under
Added Data
Rates
. Click
Delete
to remove the selected data rate from the list.
The list of procedures that depend on a data rate increases with the
corresponding number of selected custom data rates.
Nominal Data Rate
The selected
Ref. Clock Frequency
is adjusted to achieve the
Custom
Data Rate
.
Nominal Ref. Clock
For the selected
Ref. Clock Frequency
, the
Custom Data Rate
is
calculated as a multiple of the reference clock.
Ref. Clock Frequency
It can be set to 19.2 MHz, 26MHz, 38.4MHz or 52MHz.