80
Keysight N5990A MIPI M-PHY Receiver Test User Guide
3
Calibrations and Test Procedures
Sleep Time Capability
It is the sleep length required by the DUT.
Skew Tolerance
It is a limit value. If the skew between the lanes is below this limit value, the calibration process is ended.
Squelch Test Sequence
It is the pattern generator sequence for squelch tests.
Stall Length
It defines the STALL state. It specifies the number of DIFN states after the data burst. It must be specified in UI.
Stall Time Capability
It is the STALL time required by the DUT. It has a permitted range of 1 to 255 SI (1SI = 10UI).
Steps Interference Amplitude
It is the number of steps performed between the Minimum and Maximum interference amplitude points.
Step Size
This is the value to be increased/decreased at each step of the calibration/test procedure.
(HS / LS) Sync length
It defines the length of the SYNC sequence. It has a permitted range of 0 to 15 SI (1SI = 10UI).
(HS / LS) Sync range
It is the range for the SYNC sequence, It can be set as Fine or Coarse.
T Activate capability
It is the Activate time required by the DUT to exit the hibernate state and enter into the sleep state.
Target BER
The default target BER value is 1E-10 and if the measured BER is smaller than the target value, the test is
considered as passed.
Test Sequence
It is the sequence file used during the test to the DUT.
Tested Accumulated Differential
Voltage
It is the accumulated differential voltage value to be tested. The differential voltage amplitude is set
according to the calibration in order to match the desired accumulated differential voltage.
Tested Accumulated Differential
Range
It is the accumulated differential voltage range to be tested. It is defined with four values separated by
semicolons: <Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or
Logarithmic.
Tested Frequency Offset
It isthe frequency deviation relative to the nominal bit rate.
Tested Offset
It is the list of common-mode voltage values tested.
Tested Range
It is the range used to test the frequency offset value. It is defined with four values separated by semicolons:
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
Total Jitter
It is the total jitter amplitude value.
Transitions
It is the number of transitions used in each trace calibration.
Use Infiniisim
Transfer Function Chan1
Transfer Function Chan2
The parameter, “Use Infiniisim” can be set to “True” or “False”. When a replica channel is not available or you
want to use the Infiniisim Transfer Function instead of a replica channel, set the “Use InfiniiSim” parameter to
“True” and copy the Transfer Function file to the scope, either under “C:\Users\Public\Public Documents\
Infiniium\Filters\M-PHY” or “C\Documents and Settings\All Users\Documents\Infiniium\Filters\M-PHY”.
Parameter Name
Parameter Description
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......