Keysight N5990A MIPI M-PHY Receiver Test User Guide
247
PWM Tests
7
Enable Time (T
TERM-ON-PWM-RX
) Test
Table 43
Test 2.2.4-PWM-RX Differential Termination Enable Time (TTERM-ON-PWM-RX) Test
data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the value is "Fail", the termination enable time is not within the spec limit.
PWM Data Rate
It is the PWM data rate value in MBit/s.
Termination Time Min
It is the shortest/minimum termination enable time in "ns" (nanosecond).
Termination UI Min
It is the shortest/minimum termination enable time in "UI".
Min Spec
It is the minimum value of the termination enable time, where the DUT must work according to the
specification.
Termination Time Max
It is the maximum termination enable time in "ns".
Termination UI Max
It is the maximum termination enable time in "UI".
Max Spec
It is the maximum value of the termination enable time, where the DUT must work according to the
specification.
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......