3. Circuit Boards - L2100
ORTHOPANTOMOGAPH OP 2D
55
3.9 L2100 Rotation Position Sensor Board
L2100 - Location
Inside the rotating unit under the head support. To ac-
cess, remove the head holder assembly and then the
lower protection assembly (see section 7.1).
L2100 - Field replaceable parts
None.
L2100 - Description
L2100 is used to detect the position of the rotating unit.
L2100 has two optical switches that generate sensor
signals ROTSW1 and ROTSW2 according to which
position is activated.
L2100 includes a transmitter LED, and a receiver, or
base. The sensor signals remain on as long as the base
receives light from the LED. When the light to the base
is cut off by the positioning rail, the sensor signal is
switched off.
Summary of Contents for ORTHOPANTOMOGAPH OP 2D
Page 2: ......
Page 4: ...IV ORTHOPANTOMOGAPH OP 2D Contents...
Page 24: ...14 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 3 Unit dimensions...
Page 32: ...22 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Page 33: ...2 Unit Description ORTHOPANTOMOGAPH OP 2D 23 2 6 Block diagram...
Page 34: ...24 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 7 Fuse diagram...
Page 38: ...28 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Page 43: ...3 Circuit Boards L1200 ORTHOPANTOMOGAPH OP 2D 33 L1200 Block Diagram...
Page 46: ...36 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1500 L1500 Block diagram...
Page 50: ...40 ORTHOPANTOMOGAPH OP 2D 3 Circuit Board L1600 L1600 Block diagram...
Page 54: ...44 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1700 L1700 Block Diagram...
Page 57: ...3 Circuit Boards L1800 ORTHOPANTOMOGAPH OP 2D 47 L1800 Block diagram...
Page 61: ...3 Circuit Boards P1910 ORTHOPANTOMOGAPH OP 2D 51 P1910 Block diagram...