8. Aligning the unit
ORTHOPANTOMOGAPH OP 2D
113
8.6 Checking and adjusting the Chin Support
1.
PC:
Enable image capture for still mode im-
age.
Still mode can be selected using the s2termi-
nal or through the GUI of the unit.
2. Make sure that the chin support is in the 0
position and then attach the
alignment cone
assembly
to the chin support.
3. Manually rotate the Rotating Unit so that it is
parallel to the
side
of the upper shelf and the
tubehead is at the front. Insert a 4mm hexa-
gon socket wrench into the locking hole in the
guide pin assembly to hold the rotating unit in
position.