8. Aligning the unit
112
ORTHOPANTOMOGAPH
OP
2D
If the white section in the "tubehead on right"
image is
less
that the white section in the
"tubehead on left", the collimator must be
moved to the
left
(turn the collimator adjusting
screw
counter-clockwise
).
To adjust the collimator loosen the nuts (2) that
hold the collimator assembly in position and
the adjust the left/right position of the collima-
tor with the screw (1) on the right-hand side.
8. Recheck the position of the collimator and
readjust if necessary.
9. Remove the hexagon socket wrench from the
locking hole when you have completed this
adjustment.