3. Circuit Boards - CCD Sensor / Filter
ORTHOPANTOMOGAPH OP 2D
59
3.11 CCD Sensor / Filter board
CCD / Filter - Location
In rotating unit. To access, remove the sensor inner
cover (see section 7.1). Sensor
fi
lter board also re-
quires the sensor to be removed.
CCD - Field replaceable parts
None. Inside the CCD Sensor there is the CCD Sensor
board. The board cannot be accessed.
CCD - Description
CCD sensor converts X-ray radiation to visible light and
and then the CCD semiconductor chips measure the
intensity of the visible light. The analog signal is A/D-
converted and sent in parallel data lines to the CPU
board.