3. Circuit Boards - P1910
ORTHOPANTOMOGAPH OP 2D
49
3.7 P1910, PFC Board
P1910 - Location
DANGER: High voltage on this board.
Do not touch this board when the unit is switched
on.
There is high voltage on the large heat sink on this
board when the unit is switched on.
Do not touch this board until the capacitors have
discharged. After switching the unit off wait 10 min-
utes for the capacitors to discharge. Ensure also
that the indicator LED A5 is OFF.
P1910 - Field replaceable parts
Fuse FH2: SIBA FF6.3A 70 125 40. 6.3A